Transformation Efficiency Calculator

Figran

Turn colony counts into cfu/µg from one plate or a whole dilution series.

  • Several plates and dilutions
  • Check against cell specs
  • Compare transformations

DNA

100 pg transformed

Competent cells

Recovery

Recovery volume: 1,000 µL

Plates

Plate 1
Plate 2
Plate 3

Transformation efficiency

2.0 × 10⁹ cfu/µg

log₁₀ 9.29 · 95% CI 1.7–2.3 × 10⁹

196 colonies on 1 plate · DNA on selected plates: 0.1 pg

DilutionµLColoniesStatus
1:10100TNTCTNTC
1:100100196Used
1:1,00010021Below 30

Transformants in the recovery: 1.96 × 10⁵

Within the selected cell spec · NEB 5-alpha (C2987H/I/P/U) (1.0 × 10⁹–3.0 × 10⁹ cfu/µg)

Next time: 100 µL · 1:100 should give 30–300 colonies.

Transformation efficiency calculator examples

Start with an example and adapt it to your experiment.

NEB 5-alpha cells tested with pUC19

Estimate efficiency from a dilution series.

3 plates · 100 pg pUC19

NEB worked example: one plate

Calculate efficiency from the C2925 FAQ.

Source: NEB FAQ (C2925)

Two-step dilution before plating

Account for a dilution made before plating.

Source: NEB FAQ (C2992)

Electroporation efficiency of NEB 10-beta cells

Compare an electroporation result with its cell spec.

Electroporation · 10 pg pUC19

Ligation with a vector-only control

Compare a ligation with its background control.

2 transformations · NEB 5-alpha

Homemade CaCl₂ DH5α cells

Compare three batches of homemade cells.

3 batches · 1 ng pUC19

Library size from an electroporated ligation

Estimate the number of transformants in recovery.

2 plates · ligation · electroporation

pGLO teaching-lab transformation

The 510 µL total includes the DNA volume.

Source: Bio-Rad pGLO manual

How to calculate transformation efficiency

Calculate and share your result in four steps.

  1. 1Enter DNA

    Enter the DNA concentration and volume or its mass.

  2. 2Set recovery

    Choose cells and enter the recovery volumes.

  3. 3Enter plates

    Enter each dilution, plated volume and colony count.

  4. 4Take your result

    Copy the result or export your data.

Transformation efficiency formula

The transformation efficiency equation divides colony counts by DNA plated.

Efficiency
TE=coloniesDNA on plate
DNA on plate
DNA on plate=DNA transformed×VplatedVrecovery×dilution fold
Pooled efficiency
TE=∑colonies∑DNA on plate
Transformants
transformants=TE×DNA transformed

For the NEB C2992 example: 150 ÷ (0.0001 × 30 ÷ 300 ÷ 100) = 1.5 × 10⁹ cfu/µg. Exact 95% confidence intervals use the Poisson count.

References: NEB C2992 FAQ

Competent cell efficiency by strain

Compare manufacturer specifications for selected strains.

CellsChemically competent (cfu/µg)Electrocompetent (cfu/µg)Typical use
NEB 5-alphaC2987H/I/P/U1.0 × 10⁹–3.0 × 10⁹—Routine cloning
NEB 5-alphaC2987R1.0 × 10⁸–5.0 × 10⁸—Routine cloning
NEB 5-alpha F′IqC2992H/I1.0 × 10⁹–3.0 × 10⁹—Toxic gene cloning
NEB 10-betaC3019H/I1.0 × 10⁹–3.0 × 10⁹—Large plasmids
NEB 10-betaC3019P1.0 × 10⁸–3.0 × 10⁸—Large plasmids
NEB TurboC2984H/I1.0 × 10⁹–3.0 × 10⁹—Fast growth
NEB StableC3040H/I1.0 × 10⁹–3.0 × 10⁹—Unstable inserts
NEB StableC3040P1.0 × 10⁸–3.0 × 10⁸—Unstable inserts
NEB dam⁻/dcm⁻C2925H/I1.0 × 10⁶–3.0 × 10⁶—Unmethylated DNA
NEB 10-beta ElectrocompetentC3020K—> 2.0 × 10¹⁰Large plasmids
NEB Stable ElectrocompetentC3041K—> 1.0 × 10¹⁰Unstable inserts
NEB 5-alpha SubcloningC2988J> 1.0 × 10⁶—Routine cloning
Invitrogen TOP10C404010> 1.0 × 10⁹—Routine cloning
Invitrogen MAX Efficiency DH5α T1R12297016> 1.0 × 10⁹—Routine cloning
Invitrogen Subcloning Efficiency DH5α18265017> 1.0 × 10⁶—Routine cloning
Agilent XL10-Gold200315≥ 5.0 × 10⁹—Large plasmids
Promega JM109 High EfficiencyL2001> 1.0 × 10⁸—Routine cloning

These specifications are measured using control plasmid under manufacturer conditions. Electrocompetent NEB 10-beta cells have a higher specified minimum than their chemically competent counterpart.

References: NEB Competent Cell Selection Guide · Invitrogen · Agilent · Promega

Transformation efficiency calculator FAQ

What is transformation efficiency?

It is the number of colony-forming units obtained per microgram of transformed DNA, usually measured with a small amount of control plasmid. It measures bacterial transformation performance for the selected cells and DNA.

What units is transformation efficiency reported in?

It is usually reported as cfu/µg. Multiply cfu/ng by 1,000 to convert to cfu/µg.

What is a good transformation efficiency for E. coli?

Compare the result with the specification for your cells and protocol; high-efficiency and subcloning cells have different specifications.

Why is my transformation efficiency low?

Cell handling, recovery, DNA quality and antibiotic selection can affect the count. A control plasmid helps distinguish cell performance from problems with the DNA.

How do I combine plates from several dilutions?

Add the countable colonies and divide by the total DNA on those plates. Recovery volume here is cells plus medium; enter replicates as separate plates.

Which plates should I count?

Use plates in your chosen countable range when available. TNTC plates are excluded, and low counts have wider confidence intervals.

How much DNA should I use to test competent cells?

Follow your cell manufacturer’s protocol; NEB recommends clean supercoiled pUC19 in the 100 pg–1 ng range.

How do I compare a ligation with a vector-only control?

Use Compare and mark the vector-only row as background. Compare cfu/µg when DNA amounts differ; a ligation calculator can help plan the preceding reaction.

Is transformation efficiency the same as CFU/mL?

No. CFU/mL describes viable cells per volume, while cfu/µg describes transformants per mass of DNA. Transformation frequency also requires a viable-cell count. CFU Calculator.

Check your next batch of competent cells

Turn your plate counts into a result for your lab record.

Calculate efficiency