Miniprep spins at 17,900 × g
Convert several spins for a microcentrifuge.
Convert RCF to RPM and rpm to g for your rotor, then copy or print the settings.
| Step | Protocol | Set | Time |
|---|---|---|---|
| Step 1 | 12,000 × g | 11,304 rpm | — |
Eppendorf 5424 R · FA-45-24-11 · rmax 8.4 cm · no adapter
All steps within the rotor limit (15,000 rpm · 21,130 × g)
Eppendorf 5424 R · FA-45-24-11 · rmax 8.4 cm · no adapter
| × g | rpm |
|---|---|
| 100 | 1,032 |
| 200 | 1,459 |
| 300 | 1,787 |
| 400 | 2,064 |
| 500 | 2,307 |
| 1,000 | 3,263 |
| 1,500 | 3,997 |
| 2,000 | 4,615 |
| 3,000 | 5,652 |
| 4,000 | 6,526 |
| 5,000 | 7,297 |
| 10,000 | 10,319 |
| 12,000 | 11,304 |
| 14,000 | 12,210 |
| 16,000 | 13,053 |
| 20,000 | 14,593 |
| rpm | × g |
|---|---|
| 1,000 | 94 |
| 2,000 | 376 |
| 3,000 | 845 |
| 4,000 | 1,503 |
| 5,000 | 2,348 |
| 6,000 | 3,381 |
| 7,000 | 4,602 |
| 8,000 | 6,010 |
| 9,000 | 7,607 |
| 10,000 | 9,391 |
| 11,000 | 11,363 |
| 12,000 | 13,523 |
| 13,000 | 15,871 |
| 14,000 | 18,407 |
| 15,000 | 21,130 |
Start with a spin and adapt it for your rotor.
Convert several spins for a microcentrifuge.
Set a low-speed cell pellet.
Convert a non-gel blood tube separation spin.
Report relative centrifugal force from a recorded speed.
Estimate pelleting time on another rotor; acceleration is not included.
Convert a bacterial harvest speed.
The RCF calculator needs only your rotor and the protocol speed.
Select the rotor used for your spin.
Choose the tube or adapter and radius used for your force.
Enter the speed, unit and optional time for each step.
Copy or print the setup, or export a CSV and methods text.
12,000 × g · r = 8.4 cm → 11,304 rpm
Rated force uses the maximum radius. An adapter can reduce the radius and the force at a given speed.
70 min × 220 / 356 = 43 min
This estimate applies to pelleting in the same medium, not density gradients, and does not account for acceleration or braking.
References: Eppendorf 5424 R manual, §9.7.1 · Beckman Coulter, Using k-Factor
Multiply 1.118 × 10⁻⁵ by the radius in cm and rpm squared. At 14,000 rpm and 8.4 cm, the force is 18,407 × g.
It depends on the radius: 1,787 rpm at 8.4 cm, 1,264 rpm at 16.8 cm, or 1,221 rpm at 18 cm.
RCF is relative centrifugal force, expressed as a multiple of gravity in × g. RPM depends on the rotor radius.
At 8.4 cm use 10,319 rpm; at 8.6 cm use 10,198 rpm; at 7.2 cm use 11,146 rpm.
Use rₘₐₓ unless the protocol specifies another radius. Use the radius for the actual tube and adapter.
Some use 1.12 × 10⁻⁵ instead of 1.118 × 10⁻⁵, a difference of about 0.2%. Radius choice and units can cause larger differences.
For pelleting in the same medium, multiply the original time by the new k-factor divided by the original k-factor at the actual speeds. This does not apply to gradients.
Use the radius in your rotor manual or measure from the axis to the bottom of the tube. Select Custom radius to enter it.